medical Class II Updated 2020-06-17

RAYSEARCH LABORATORIES AB recalls RayStation 8B Service Pack 2, UDI # 07350002010235 - Produc

Recalled Product

RayStation 8B Service Pack 2, UDI # 07350002010235 - Product Usage: For radiation therapy and medical oncology which can also be used to administer treatments.


Hazard / Issue

Software error was identified in the Pencil Beam Scanning (PBS) and Line Scanning (LS) dose calculation algorithms that could result in a local underestimation of expected dose.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI # 07350002010235 Build number: 8.1.2.5