medical Class II Updated 2019-08-28

Siemens Medical Solutions USA, Inc. recalls ACUSON NX2 Elite Diagnostic Ultrasound System, Model # 11284

Recalled Product

ACUSON NX2 Elite Diagnostic Ultrasound System, Model # 11284500, RX Only, Siemens Medical Solutions, containing software version VA10A, VA10B, VA11A, VA11B; UDI # 04056869050775 The ACUSON NX2 diagnostic ultrasound imaging systems are intended for the following applications: Cardiac, Fetal, Abdominal (including liver), Pediatric, Small Parts (Small Organ), Transcranial, OB/GYN, Pelvic, Urology, Vascular (including Peripheral Vessel) and Musculoskeletal applications. The systems also provide for the measurement of anatomical structures and for analysis packages that provide information that is used for clinical diagnosis purposes.


Hazard / Issue

The action is being initiated due to internal testing which identified a possibility for transducers to exceed the acoustic output power (AOP) values defined . The transit voltage values based on PW Doppler are calculated normally but not loaded correctly into hardware

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All product lots of the Acuson NX2, model # 11284500 containing the following software: VA10A, VA10B,VA10C, VA11A, VA11B, VA11C
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →