medical Class II Updated 2023-08-16

Waldemar Link GmbH & Co. KG (Mfg Site) recalls LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 21

Recalled Product

LINK MP Monoblock Hip Stem, Trial Stem Instruments, Size 21 - Indicated for patients with Mobility-limiting diseases, fractures or defects of the hip joint or proximal femur Item Number: 136-121/00


Hazard / Issue

Increased force may be required to remove the LINK MP Monoblock trial hip stem (Sizes 14mm - 25mm),after impaction. Potential for prolongation of surgery or the surgical procedure may have to be modified due to this issue

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 04026575182152 Lot Numbers: B922120 B933167 C005021 C211015
View official government recall