medical Class II Updated 2023-08-16

Siemens Medical Solutions USA, Inc recalls ARTIS icono floor-Angiography systems developed for single a

Recalled Product

ARTIS icono floor-Angiography systems developed for single and biplane diagnostic imaging and interventional procedures. Material Number: 11327700


Hazard / Issue

Due to a hardware issue in the cable connectors of the system generator, a thermal overload in the cable connections may occur when performing excessive fluoroscopy/acquisitions, a burning smell may be detectable coming from the generator cabinet and the system may lose the imaging functionality of the corresponding plane and result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: 4056869149325 Serial Numbers: 170561 170549 170559 170546 170552 170571 170570