medical Class II Updated 2023-08-16

Randox Laboratories Ltd. recalls Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers

Recalled Product

Randox RX Series Urea Enzymatic Kinetic Assay Ref. Numbers: (1) UR3825 (2) UR3873 (3) UR8334 (4) UR8070 (Modena)


Hazard / Issue

Carryover from the Direct LDL-cholesterol assay on the Rx instruments may result in a negative bias, up to -11%, in quality control and patient samples, in urea assays results if the urea assay is run immediately after the Direct LDL-cholesterol assay. This may lead to a delay in reporting results or in falsely decreased urea test results.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Catalogue Number/GTIN (1) UR3825 05055273206906; (2) UR3873 05055273206913 ; (3) UR8334 05055273209600; (4) UR8070 05055273209594. Not Batch Specific.