medical MODERATE Updated 2026-06-17

The Binding Site Group, Ltd. recalls Brand Name: EXENT Analyser Product Name: EXENT Analyser Mo

See all recalls from The Binding Site Group, Ltd. →

Recalled Product

Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Brand Name: EXENT Analyser →
View official government recall