medical Class I Updated 2023-08-23

Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Cat

Recalled Product

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 16 cm, Curved Extensions, IC Tray REF: 8888101002HP


Hazard / Issue

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GTIN: 20884521127965 LOT Numbers: 2133600171 2133600172 2133700047 2133700048 2133700060 2200400110 2204000236 2230400213 2230400214 2233400068
View official government recall