medical Class I Updated 2023-08-23

Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Cat

Recalled Product

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 20 cm, Curved Extensions, IC Tray REF: 8888101003HP


Hazard / Issue

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GTIN: 20884521127972 LOT Numbers: 2130900104 2133600083 2133600160 2133600165 2226500171 2226500173 2227800101 2227800104 2227800109 2230400208