medical Class I Updated 2023-08-23

Covidien LP recalls MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Cat

Recalled Product

MAHURKAR 12 Fr High Pressure Triple Lumen Acute Dialysis Catheter, 24 cm, Straight Extensions, Kit REF: 8888340637HP


Hazard / Issue

Catheter center lumen was found to have an occlusion in the tip of the catheter; the source of the occlusion an excessive MDX, a silicone-based lubricant which coats the catheter tip. An occluded, partially or fully, and/or uncured or excessive MDX may dislodge from the catheter lead to full catheter obstruction resulting in delay to treatment and partial obstruction resulting in reduced flow, or particulate dislodgement that may result in delay to treatment, hemolysis, embolism/embolus or thrombosis/thrombus

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GTIN: 10884521128088, 20884521128085 LOT Numbers: Lot Number 1907700088 2015100039 2019500217 2107700128 2113300287