medical Class II Updated 2021-09-08

Aspen Surgical Products, Inc. recalls Norton Arm Kit, Product Code 711400, (previous product code

Recalled Product

Norton Arm Kit, Product Code 711400, (previous product code 031001), Sample Kit S711400


Hazard / Issue

Affected products were manufactured with materials found to have a bioburden level exceeding the limit at which the product's sterilization process was validated. As a result, the firm cannot provide full assurance that the affected products were effectively sterilized, and the product may compromise a sterile field and increase the risk of a surgical site infection.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots manufactured between July 9, 2019 to July 9, 2021