medical Class II Updated 2023-08-23

Olympus Corporation of the Americas recalls UroPass Ureteral Access Sheaths, 5 pieces/box

Recalled Product

UroPass Ureteral Access Sheaths, 5 pieces/box


Hazard / Issue

Dilator tips may break in the package and in patients during surgical procedures.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Models: 61024BX, 61038BX, 61046BX, 61054BX, 61124BX, 61138BX, 61146BX, 61154BX, 61224BX, 61238BX, 61246BX, 61254BX, 61324BX, 61338BX, 61346BX, 61354BX Lots: All products manufactured in 2018 and 2019. Manufacturing date is printed on pouch label in YYYY-MM-DD format.