medical Class II Updated 2018-07-18

Stryker GmbH recalls Femoral Nail, LEFT T2 GTN ¿10x360 mm, Product Number 1850103

Recalled Product

Femoral Nail, LEFT T2 GTN ¿10x360 mm, Product Number 18501036S


Hazard / Issue

The manufacturer has discovered that potentially out-of-specification products may have left the factory. Non-conforming cannulation of the nails may result in reduced component strength and potentially premature nail breakage.

Issued by

FDA

Affected States: AL, FL, KY, NC, TN, UT, WA
Lot/Code Info: GTIN: 04546540646583; Lot code: K08D58D