medical Class I Updated 2020-07-08

Arrow International Inc recalls Arrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE,

Recalled Product

Arrow AutoCAT2 Intra-Aortic Balloon Pump AEROAUTOCAT 2 WAVE, Product Code IAP-0535


Hazard / Issue

A potential issue with a component within the above-referenced IABPs may impact the ability of the devices to operate. A Component within the IABP is susceptible to vibration failure resulting in fretting, charring, and discoloration of motor connector wires, which may result in pump alarms for System Error 3 and High Baseline presented on the screen of the IABP, and potential abrupt cessation of function or inability to start the IABP.

Issued by

FDA

Distribution: Worldwide distribution.
Lot/Code Info: Serial Number Range: AEROAUTOCAT 2 WAVE IAP-0535 150907W 151030W 151101W 151104W 151106W 160305W through 160306W 160743W 160824W 190119W through 190120W 190221W 190223W through 190226W 191047W
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →