medical Class II Updated 2018-07-18

Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF) recalls Medtronic EnTrust SingleChamber Implantable Cardioverter Def

Recalled Product

Medtronic EnTrust SingleChamber Implantable Cardioverter Defibrillator with Atrial and Ventricular Therapies and RapidRead Telemetry, D153VRC Product Usage: The implantable cardioverter defibrillator is intended to provide ventricular antitachycardia pacing and ventricular defibrillation for automated treatment of life-threatening ventricular arrhythmias.


Hazard / Issue

Medtronic has become aware of a potential for loss of high voltage and anti-tachycardia pacing therapy in EnTrust and Escudo implantable cardioverter defibrillators (ICDs) as they near elective replacement indicator (ERI) voltage. Under certain circumstances, the device may display an immediate End of Life (EOL) Observation with no prior ERI alert. Though no ERI alert is triggered, there may not be enough remaining battery capacity to charge the high voltage circuits, resulting in an excessive charge time EOL Observation, leading to a loss of high voltage and anti-tachycardia pacing therapy.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Product UPN/GTIN: 00885074300245 Serial Numbers: PNT402938H PNT413911H PNT413995H