medical Class II Updated 2020-07-08

Westmed, Inc. recalls Neo-Vent Infant T-Piece Resuscitation Circuit;Circuit for us

Recalled Product

Neo-Vent Infant T-Piece Resuscitation Circuit;Circuit for use on compatible Infant Resuscitators. 15mm OD x 10mm ID connector; REF 0415, Qty:10, Length: 60" (1.5cm), Rx Only.


Hazard / Issue

A repair was made to an adapter mold resulting in parts being produced with compliant 10mm ID connections and non-compliant 15mm OD connections. This resulted in facilities with infant resuscitators that connect via the 15mm OD fitting not being able to use those circuits.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: 101118T42, 101118T44, 102918T01, 080618T02, 080618T03 , 101118T43, 102918T02.
View official government recall