medical Class II Updated 2019-09-11

Integra LifeSciences Corp. recalls Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and

Recalled Product

Integra Jarit GEMINI CLAMP 9 1/4 inch - used in vascular and cardiothoracic procedures used for clamping and occluding vessels. Cat. #140-332


Hazard / Issue

A small crack at the box lock corner may after multiple cleaning and sterilization cycles produce an increased risk of corrosion, weakening the instrument and resulting in premature need for replacement and/or breakage of the instrument

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot #AA1901