medical Class II Updated 2020-07-08

Mediana Co., Ltd. recalls DIRECT SUPPLY Attendant VITAL SIGNS MONITOR, REF AVSM2 - Pro

Recalled Product

DIRECT SUPPLY Attendant VITAL SIGNS MONITOR, REF AVSM2 - Product Usage: It is vital sign monitor to monitor noninvasive blood pressure (NIBP), functional arterial oxygen saturation (SpO2), pulse rate (PR), temperature (Temp) for adult, pediatric and neonatal patients in all areas of a hospital and hospital-type facilities.


Hazard / Issue

Failures were found in the temperature measurement function while in the predict mode.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 456802000001~456802000180
View official government recall