medical Class II Updated 2020-07-08

Radiometer Medical ApS recalls ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UD

Recalled Product

ABL80-FLEX CO-OX 393-841 Box Label, REF 393-841, IVD, CE, UDI: 05700693938417 - Product Usage: It is intended for use in a laboratory environment, near patient or point-of-care setting.


Hazard / Issue

Barcode readers on ALB900 and ABL800 analyzers used for measuring pH, blood gases, electrolytes, glucose, lactate, hematocrit and oximetry may misinterpret the contents of some barcode types not using a check digit for patient ID or accession number. Misinterpretation of a barcode, including e.g. patient ID or accession number, may result in patient mix up and/or delayed medical treatment.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: All lots
View official government recall