medical Class II Updated 2018-07-25

Ion Beam Applications S.A. recalls Proteus 235, Universal Nozzle with snout 300x400; PBS dedica

Recalled Product

Proteus 235, Universal Nozzle with snout 300x400; PBS dedicated nozzle with snout XL Product Usage: Device is designed to produce and deliver proton beam for the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation. The PTS may include a fixed small beam treatment room dedicated to the treatment of patients with localized tumors and other conditions susceptible to treatment by radiation localized to the head and neck.


Hazard / Issue

Gantry rotation or proton irradiation can be allowed even if apertures or accessories are not correctly positioned in snout 300x400 or snout XL. This can lead to risk of crushing for patient and/or user, and risk of mistreatment for the patient.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: PAT.003, PAT.006, PAT.107, PAT.108, PAT.109, PAT.110, PAT.111, PAT.113, PAT.114, PAT.115, PAT.116, PAT.120, PAT.126, and PAT.132.