Siemens Healthcare Diagnostics, Inc. recalls Atellica CH Total Bilirubin_2 -In vitro diagnostic use in th
Recalled Product
Atellica CH Total Bilirubin_2 -In vitro diagnostic use in the quantitative determination of total bilirubin in human serum and plasma (lithium heparin) of adults and neonates Test Code: TBil_2 Siemens Material Number (SMN): 11097531
Hazard / Issue
Potential for reagent carryover on the Atellica CH 930 resulting in a positive bias that could impact quality control (QC), patient samples, and calibrator results with from Atellica¿ CH Total Bilirubin_2 (TBil_2), Atellica¿CH LDL Cholesterol (LDLC), Atellica¿ CH Gamma-Glutamyl Transferase (GGT) and Atellica¿CH HDL Cholesterol (HDLC)