Philips North America Llc recalls The Allura Xper series is intended for use on human patients
Recalled Product
The Allura Xper series is intended for use on human patients to perform: "Vascular, cardiovascular and neurovascular imaging applications, including diagnostic, interventional and minimally invasive procedures. This includes, e.g., peripheral, cerebral, thoracic and abdominal angiography, as well as PTAs, stent placements, embolisations and thrombolysis. "Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA, stent placing, atherectomies), pacemaker implanatations, and electrophysiology (EP). "Non-vascular interventions such as drainages, biopsies and vertebroplasties procedures."
Hazard / Issue
Early life failure of the ADUs only occurs when the hospital mains provide 480V to the system. This current voltage may lead to saturation and overheating of the coils that protect the IGBTs (insulated-gate bipolar transistor) of the ADU. The overheating may generate a peculiar burning smell that may be noticed. In some instances, this burning smell may be noticed before the ADU fails. When the ADU fails the X-ray performance of the system is reduced to Emergency Fluoroscopy. Exposure will not be possible and image quality will be reduced.
Issued by
FDA
Were you affected by this recall?
If you were harmed by a recalled product, you may have legal options. Consider speaking with a consumer protection lawyer in your state.
Looking for safe alternatives?
Browse certified-safe products in this category on Amazon →
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.
CAN-SPAM compliant · No spam · Unsubscribe anytime