medical Class II Updated 2018-07-25

Philips Medical Systems (Cleveland) Inc recalls Ingenuity Digital PET/CT System, Model 882446 with the follo

Recalled Product

Ingenuity Digital PET/CT System, Model 882446 with the following UDI's: (01)00884838059559(21)900004 (01)00884838059559(21)900005 (01)00884838059559(21)900006 (01)00884838059559(21)900007 (01)00884838059559(21)900008 (01)00884838059559(21)900009 (01)00884838059559(21)900010 (01)00884838059559(21)900011 (01)00884838059559(21)900012 (01)00884838059559(21)900013 (01)00884838059559(21)900014 (01)00884838059559(21)900015 (01)00884838059559(21)900016 (01)00884838059559(21)900017 (01)00884838059559(21)900018 (01)00884838059559(21)900019 (01)00884838059559(21)900020 (01)00884838059559(21)900021 (01)00884838059559(21)900022 (01)00884838059559(21)900023 (01)00884838059559(21)900024 (01)00884838059559(21)900025 (01)00884838059559(21)900026 (01)00884838059559(21)900027 (01)00884838059559(21)900028 (01)00884838059559(21)900029 (01)00884838059559(21)900030 (01)00884838059559(21)900031 (01)00884838059559(21)900032 (01)00884838059559(21)900033 (01)00884838059559(21)900034 (01)00884838059559(21)900035 Product Usage: The system is intended to image the whole body, heart, brain, lung, gastrointestinal, bone, lymphatic, and other major organs for a wide range of patient types, sizes, and extent of diseases. Both subsystems can also be operated as fully functional, independent diagnostic tools, including application of the CT scanner for diagnosis and for use in radiation therapy planning.


Hazard / Issue

The block used to connect the patient pallet to the couch may be assembled incorrectly. A screw may be misaligned in the patient pallet.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 900004 900005 900006 900007 900008 900009 900010 900011 900012 900013 900014 900015 900016 900017 900018 900019 900020 900021 900022 900023 900024 900025 900026 900027 900028 900029 900030 900031 900032 900033 900034 900035
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →