medical Class II Updated 2018-08-01

Philips Healthcare recalls Centron Product Usage: Vascular,cardiovascular and neu

Recalled Product

Centron Product Usage: Vascular,cardiovascular and neurovascular imaging applications,including diagnostic, interventional and minimally invasive procedures. Cardiac imaging applications including diagnostics, interventional and minimally invasive procedures (such as PTCA,stent placing,atherectomies), pacemaker implantat ions, and electrophysiology (EP). Non-vascular interventions such as drainages,biopsies and vertebroplasties procedures.


Hazard / Issue

The first time an operator selects a new procedure type during a single examination,the shutter position resets to the open position for the new procedure type. If the shutters had previously been changed during the examination, that setting is not retained after the first time the procedure type is changed during a single examination.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Centron 1.0.10 Centron 1.0.10.1 Centron 1.0.10.5
View official government recall