COVIDIEN MEDTRONIC recalls Tri-Staple 2.0 Black Intelligent Reload for use with Signia

Recalled Product

Tri-Staple 2.0 Black Intelligent Reload for use with Signia stapling system 60 mm Extra Thick, Item Code SIG60AXT


Hazard / Issue

Device may be missing a sled component. The sled component is responsible for staple deployment. This may result in failure to form a staple line when tissue is divided, leading to bleeding or leakage of luminal contents.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: N7L0852X, N7M0130X, N8A0145X, N8A0336X, N7L0931X, N7M0639X, N8A0796X & N8B0993X