Reichert, Inc. recalls PHOROPTOR VRX DIGITAL REFRACTION SYSTEM-Digital refractor in
Recalled Product
PHOROPTOR VRX DIGITAL REFRACTION SYSTEM-Digital refractor intended for the subjective measurement of the refractive error of the eye. MODEL NUMBER: 16242 [Model 16241 is the base model. Model 16242 includes an additional prism feature used for the measurement of ocular deviation in patients with latent or manifest strabismus].
Hazard / Issue
Phoroptor VRx Head may detach from the bracket that connects the Phoroptor VRx Head to an ophthalmic stand. If the Phoroptor VRx Head detaches from the bracket, it can fall onto a patient or device operator