medical MODERATE Updated 2026-07-01

Ventec Life Systems, Inc. recalls VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF:

See all recalls from Ventec Life Systems, Inc. →

Recalled Product

VOCSN+Pro package, REF: PRT-00853-000; VOCSN+Pro unit, REF: PRT-00490-001

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Respiratory system intended to provide continuous/intermittent ventilator support, may have a damaged back-up battery supporting backup alarm function, which may result in the backup alarm not activating during loss of primary power and absence/depletion of backup power sources, which may result in delayed caregiver awareness of a hazardous condition.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: REF/UDI-DI/Serial Number/Lot: PRT-00853-000/00855573007839/155391P/807138, PRT-00490-001/00855573007792/155391/804613