Siemens Medical Solutions USA, Inc recalls Artis zee systems with Laird Cooling Unit for SSFD: Material

Recalled Product

Artis zee systems with Laird Cooling Unit for SSFD: Material # 7555118. Artis is an angiography system developed for single and biplane diagnostic imaging and interventional procedures.


Hazard / Issue

Due to a defective sealing, coolant may inflow into the electrical parts of the system cabinet, which could cause a short circuit and cause the system to shut down suddenly during an ongoing procedure

Issued by

FDA

Affected States: HI, IL, MI, NC, NY, VA, WI
Lot/Code Info: Laird Cooling Unit for SSFD: Material # 7555118.