medical MODERATE Updated 2026-07-01

PFM MEDICAL INC. recalls Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm,

See all recalls from PFM MEDICAL INC. →

Recalled Product

Multi-Snare Set: 5 mm x 125 cm, REF: 147305; 10 mm x 125 cm, REF: 147310

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

A snare catheter including luer lock, indicated for removal or manipulation of a foreign body in the vascular system, failed to meet the biocompatibility requirements for in-vitro-cytotoxicity, which if in contact with a patient's vasculature could result in localized inflammation and tissue irritation.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: REF/UDI-DI/Lot(Expiration): 147305/04042301061829/1049868(10/14/2028); 147310/04042301033567/1049960(10/20/28)

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Multi-Snare Set: 5 mm →
View official government recall