medical Class II Updated 2020-07-15

Bound Tree Medical recalls Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tub

Recalled Product

Curaplex Ambu King LTS-D Supraglottic Airway Kit with ET Tube Part Number: KLTSD404K Curaplex Ambu¿ King LTS-D" Supraglottic Airway Kit with ET Tube (hereafter the Kit which contains all of the components necessary to successfully obtain an open patient airway with an Ambu¿ King LTS-D" Supraglottic Airway.


Hazard / Issue

Kit contained an incorrect catheter-tip syringe not compatible with the luer lock connector of the included airway. As a result, may be unable to inflate the balloons of the supraglottic airway to create a secure seal in the patient s airway as intended by the kit

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot Number: ASM0025699
View official government recall