Ad-Tech Medical Instrument Corporation recalls AD-TECH Medical Instrument Corporation Drill Sleeve Guides D
Recalled Product
AD-TECH Medical Instrument Corporation Drill Sleeve Guides DSG-6-3-090-2.4N utilizing raw material (RM0677) from the following lots from vendor 00-0001062., AD0021502, AD0021628. Product Usage: The drill sleeve guides are accessories to the Ad-Tech Cranial Drill Bits. The Cranial Drill Bits and accessories are intended to be used for drilling holes in the skull for neurological procedures, such as brain biopsy, brain contacting electrode and electrode accessory device placement. The DSG-6.3-090-2.4N drill sleeve guides are intended to be used with the 2.4mm Cranial Drill Bit diameter.
Hazard / Issue
The diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Issued by
FDA
Were you affected by this recall?