Ad-Tech Medical Instrument Corporation recalls AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kit
Recalled Product
AD-TECH Drill Sleeve Guides, used with Ad-Tech Placement Kits: (a) REF DSG-90-2.3N (b) REF DSG-6.3-090-2.4N
Hazard / Issue
The inner diameter of the drill sleeve guide raw material was found to be under tolerance, potentially resulting in the drill bit seizing in the guide during surgery.
Issued by
FDA
Affected States: NY, PA, WI, AZ, IL, CA, NH, GA, MA, MN, UT, MI, WA, FL, TX, OH, OR, MD
Lot/Code Info: All serial numbers distributed between January 2014 and May 2019