medical Class II Updated 2019-10-02

GETINGE US SALES LLC recalls The HEMASHIELD PLATINUM Woven Double Velour Vascular Graft i

Recalled Product

The HEMASHIELD PLATINUM Woven Double Velour Vascular Graft is a woven double velour polyester graft impregnated with highly purified bovine collagen. The HEMASHIELD PLATINUM Woven Double Velour Vascular Graft is designed to reduce bleeding at implant and thereby eliminates the operative preclotting step, including cumbersome autoclave techniques. The collagen is designed to be gradually resorbed by the patient. The CONCENTRICRIMP pleat and GUIDELINE stripe are also featured. In addition to collagen, the graft also contains glycerol as a softening agent.


Hazard / Issue

Based on an internal incident report, Getinge/Intervascular SAS has identified that the one (1) HEMASHIELD GOLD Knitted Microvel Double Velour Vascular Bifurcated Graft product with the diameter size of 16 x 9 mm is likely incorrectly packed inside of the carton which was labeled to contain a HEMASHIELD PLATINUM Woven Double Velour Vascular Graft product with the diameter size of 16 x 8 mm ((REF: M00202166168P0, Lot: 18K31, SN: 1190443415, size 16 x 8 mm).

Issued by

FDA

Affected States: NJ
Lot/Code Info: Lot Code: 18K31 Serial Number: 1190443415 Part Number: M00202166168P0 UDI Code: 00384401018452