medical Class II Updated 2018-08-08

Randox Laboratories Ltd. recalls Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product

Recalled Product

Liquid Cardiac Control - Level 3, cat. no. CQ5053 Product Usage: This product is intended for in vitro diagnostic use, in the quality control of Cardiac Markers on clinical chemistry and Immunoassay systems.


Hazard / Issue

The device is not suitable for the control of the Troponin T assay due to unacceptable variation between vials. The deviation in precision observed is up to 28%. IQC that is reported as out of range could lead to a delay in reporting Troponin T results. If the failure occurs the operator may observe the quality control result for Troponin T falling outside range and/or increased imprecision for Quality Control monitoring.

Issued by

FDA

Affected States: WV
Lot/Code Info: Lot 4245CK.