medical Class II Updated 2023-09-20

PTW-FREIBURG recalls BeamAdjust software Version 2.3.3 Ref: S080032 The follow

Recalled Product

BeamAdjust software Version 2.3.3 Ref: S080032 The following devices are affected when a relative calibration is performed using ArrayCal (module from BeamAdjust S080032, UDI-DI: EPTWS0800320 ): " OCTAVIUS Detector 1600 SRS (T10056) (UDI-DI: EPTWT100560) " OCTAVIUS Detector 1600 MR (T10057) (UDI-DI: EPTWT100570) " OCTAVIUS Detector 1600 XDR (T10058) (UDI-DI: EPTWT100580) " OCTAVIUS Detector 1000 SRS (T10036) (UDI-DI: EPTWT100360)


Hazard / Issue

Generated calibration file will not be correct in case of a relative calibration of an OCTAVIUS Detector 1000 or OCTAVIUS Detector 1600, if a Tiff file or a DICOM data set is used as reference matrix. Therefore, the measurement result will be incorrect by using such a calibration file.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI: EPTWS0800320 BeamAdjust All software Versions
View official government recall

Looking for safe alternatives?

Browse certified-safe products in this category on Amazon →