medical Class II Updated 2018-08-08

GE Healthcare, LLC recalls Discovery IQ 2.0 20cm, 4 ring, Model Number 5432539-22

Recalled Product

Discovery IQ 2.0 20cm, 4 ring, Model Number 5432539-22


Hazard / Issue

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number 5432539-22. Serial Number 502674HM5 (System ID Number 510204DISCIQ); Serial Number 504305HM4 (System ID Number 219922PTIQ)