GE Healthcare, LLC recalls Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Dis
Recalled Product
Discovery MI, labeled as Discovery DMI 15 cm 3 ring, and Discovery DMI 20cm 4 ring
Hazard / Issue
Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).
Issued by
FDA
Distribution: Nationwide (United States)
Lot/Code Info: Model Number 5454001-170. Serial Number (System ID Number): CJRPX1800016CN ( 4550966PT11), CJRPX1800017CN (561422MI), CJRPX1800018CN (617726PET1EB), CJRPX1800021CN (404616DMI)