Biodex Medical Systems, Inc. recalls AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through Revis
Recalled Product
AtomLab 500Plus Dose CalibratorSoftware 2.0.00 through Revision 2.0.08, inclusive Model: 086-332
Hazard / Issue
When deleting a previously entered custom isotope, the software deletes the isotope, but not the associated dial setting. Possible other stored custom isotopes may then have incorrect dial settings and may lead to an incorrect dose measurement and/or misadministration