medical Class II Updated 2018-08-08

GE Healthcare, LLC recalls Revolution CT 160 1.5D STD WAUK

Recalled Product

Revolution CT 160 1.5D STD WAUK


Hazard / Issue

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Model Number 5590000-6. Serial number REVVX1800021CN (System ID 812842REV, UDI 01008406821185521118020021REVVX1800021CN), Serial Number REVVX1800004CN (System ID 269375REVOCT, UDI 01008406821185521118010021REVVX1800004CN).
View official government recall