medical Class II Updated 2018-08-08

GE Healthcare, LLC recalls Optima CT 520

Recalled Product

Optima CT 520


Hazard / Issue

Some CT and CT/PET systems recently installed with a GE supplied A! electrical panel may not be properly wired to the partial system UPS (Uninteruptible Power Supply).

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Serial number, CBCRG1700051HM, System ID Number 580622O520, UDI Number 01008406821025681117120021CBCRG1700051HM