medical MODERATE Updated 2026-07-08

Medline Industries, LP recalls Medline Convenience kits containing recalled Swan-Ganz Cathe

See all recalls from Medline Industries, LP →

Recalled Product

Medline Convenience kits containing recalled Swan-Ganz Catheters RIGHT HEART CATH PACK CTX DYNJ68530B

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI-DI 10195327345044 Lots 23DBI807 23FBG708 23GBK642 23KBO477 24BBG198 24FBN889 24GBJ879 24HBR045 24LBT492 25ABL957 25BBN572 25GBT734 25HBQ817 25LBM617 26BBK766 26DBE720

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
View official government recall