medical MODERATE Updated 2026-07-08

Medline Industries, LP recalls Medline Convenience kits containing recalled Swan-Ganz Cathe

See all recalls from Medline Industries, LP →

Recalled Product

Medline Convenience kits containing recalled Swan-Ganz Catheters OPEN HEART DYNJ909090B OPEN HEART CDS QVH CDS982523M

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Kits contain Swan-Ganz catheters recalled by Becton Dickinson (BD). BD has received complaints of leaks/breakage involving failures of the (blue) proximal injectate lumen hub on these catheters. This has led to instances of leakage, lumen damage, and breakage, and may lead to infection, medication loss, and/or blood loss.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: DYNJ909090B UDI-DI 10195327652432 Lots 24EBL051 24GBG271 CDS982523M UDI-DI 10195327560591 Lots 24ALA227 24BLA369 24CLA108 24CLA518 24DLA082 24ELB053 24ILA112

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
View official government recall