medical Class II Updated 2023-09-27

Remote Diagnostic Technologies Ltd. recalls Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-

Recalled Product

Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, and 00-1007, when Used With Tempus Pro Video Laryngoscope, REF 01-2044


Hazard / Issue

Full screen error message may occur on patient monitors with affected hardware version, either during or after laryngoscopes are unplugged from monitors, informing users a 60-100 second shutdown/restart are required. Error prevents: viewing graphical patient vital signs, sounding of pulse audio tone, visualization of patient airway used during intubation; may lead to treatment delay or hypoxia.

Issued by

FDA

Affected States: AZ, AL, CA, CO, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NM, NV, NY, OH, OK, OR, PA, TN, TX, UT, VA, WA, WI, WV, WY
Lot/Code Info: Monitors with hardware version: Trizeps-7, and software versions v07.26 through v07.30. REF/UDI-DI: 00-1004-R/05060472440020, 05060472442901; 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932; 00-1004/05060472440624; 00-1007/05060472440655
View official government recall