Fresenius Kabi USA, LLC recalls Ivenix Infusion System (IIS), Large Volume Pump (LVP) Softwa
Recalled Product
Ivenix Infusion System (IIS), Large Volume Pump (LVP) Software: 5.0.0 and 5.0.1 Part Number: LVP-SW-0004. Indicated for use in a hospital and in outpatient care environments for the controlled administration of fluids through clinically accepted routes of administration.
Hazard / Issue
Start-up time for the flow rate range greater than 50 ml/hr, and less than 200 ml/hr may exceed the specified 6 minutes, depending on the characteristics of air bubble entrapment within the Intermediate Pumping Chamber (IPC) of the administration set
Issued by
FDA