medical Class II Updated 2023-10-04

3M Company - Health Care Business recalls 3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF

Recalled Product

3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air removal steam sterilization processes


Hazard / Issue

Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132¿C and 135¿C.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: UDI/DI: 50707387787276( (shipper); Lot Codes: 33P3EF, exp. 12/27/2024; 33P4ND, exp. 01/01/2025; 33P4NR, exp. 01/01/2025; 33P5CA, exp. 01/03/2025; 33P5XT, exp. 01/06/2025