Baxter Healthcare Corporation recalls Baxter Epiphany Cardio Server E3 ECG Management System Serve
Recalled Product
Baxter Epiphany Cardio Server E3 ECG Management System Servers with software versions: a) v6.1.x, b) v6.2.x, c) v7.0.x.
Hazard / Issue
Epiphany, now a Baxter Healthcare company, is issuing an Urgent Medical Device Correction for the Cardio Server E3 ECG Management System software listed below due to a report of the caliper tool becoming non-responsive, preventing the caliper function from being used temporarily on software versions v6.1.x, v6.2.x, and v7.0.x of the E3 user interface.