medical Class II Updated 2023-10-11

Qiagen GmbH recalls therascreen ¿KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay

Recalled Product

therascreen ¿KRAS RGQ PCR Kit (24)-IVD qualitative PCR assay used on the Rotor-Gene Q MDx instrument for the detection of seven somatic mutations in the human KRAS oncogene REF 870021 (US IVD)


Hazard / Issue

Handbook Revision (Rev13) a dilution step is indicated as an Action when having the Flag SAMPLE_INT_CTRL_FAIL. A dilution of CRC tissue could lead to a false negative result. To remove ambiguity, an alteration must be made to the S the sample result with this flag should be interpreted as invalid

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: GTIN number: 04053228002048 All lots that have an expiry date of March 2023 or later
View official government recall