medical Class II Updated 2020-07-29

Medtronic Sofamor Danek USA Inc recalls Accurian RF Ablation,Cannula pk, Accurian, 100mm, 22G, 10mm,

Recalled Product

Accurian RF Ablation,Cannula pk, Accurian, 100mm, 22G, 10mm, SS Product Number: AC0019 GTIN / UPN: 20763000153667 - Product Usage: (sterile, single-use) comprises an electrically insulated shaft, an active electrode for RF energy delivery, a hub with a female slip luer fitting, and a stylet with a color-coded cap to indicate gauge size. The ACCURIAN RF Cannula included in the ACCURIAN RF Ablation Platform intended for the creation of radiofrequency lesions in nervous tissue.


Hazard / Issue

Insertion difficulty of the probes. The probe was catching on a lip inside the hub of the cannula.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lot # BY19D020
View official government recall