medical HIGH RISK Updated 2017-07-26

Penumbra Inc. recalls Penumbra 3D Revascularization Device It is indicated for us

See all recalls from Penumbra Inc. →

Recalled Product

Penumbra 3D Revascularization Device It is indicated for used in the revascularization of patients with acute ischemic stroke secondary to intracranial large vessel occlusive disease (within the internal carotid, middle cerebral-M1 and M2 segments) within 8 hours of symptom onset. Patients who are ineligible for intravenous tissue plasminogen activator (IV t-PA) or who fail IV t-PA therapy are candidates for treatment.

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Penumbra has identified an issue in these four lots involving a raw material component of the delivery wire. This issue could result in breakage of the delivery wire, which could potentially lead to serious patient injury or death.

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Lots C00644, C00645, C00646, C00717