Daavlin Distributing Company recalls Photera 600, phototherapy device; Ultraviolet Lamp For Derma
See all recalls from Daavlin Distributing Company →Recalled Product
Photera 600, phototherapy device; Ultraviolet Lamp For Dermatologic Disorders
Hazard / Issue
Calibrated Output was not calculating correctly - UV Sensor GAIN being applied was incorrect - GAIN seemed to be "4/5" instead of "7" (counts were at ~1000 instead of ~5000).
Issued by
FDA
Distribution: Nationwide (United States)
Lot/Code Info: Serial Numbers: 6-09033NB4CL, 6-09037NB4CL, 6-09031NB4CL, 6-09538NB4CL, 6-09038NB4CL, 6-09039NB4CL, 6-09040NB4CL, 6-09041NB4CL, 6-09035NB4CL, 6-09034NB4CL, 6-09030NB4CL, 6-09026NB4CL, 6-09023NB4CL, 6-09028NB4CL, 6-09025NB4CL, 6-09043NB4CL, 6-09029NB4CL, 6-09044NB4CL, 6-09022NB4CL, 6-09027NB4CL, 6-09036NB4CL, 6-09045NB4CL, 6-09024NB4CL, 6-07827NB4CL, 6-09539NB4CL, 6-09042NB4CL.
✅ What Should You Do?
- 1Contact your healthcare provider immediately to discuss this recall.
- 2Stop using the device if it is safe to do so, and contact your provider.
- 3Follow the manufacturer's remedy instructions provided in the recall notice.
- 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.
Looking for a safe alternative?
Search Amazon for Photera 600, phototherapy device; →If you were harmed by this recalled product, Top Class Actions lists active class action lawsuits and settlements that may apply to you.
Get the next US recall alert before it hits the news.
Free. Weekly digest or instant alerts. Unsubscribe anytime.
CAN-SPAM compliant · No spam · Unsubscribe anytime