medical MODERATE Updated 2026-08-05

Terex Utilities recalls Power Over Fiber Base Module

See all recalls from Terex Utilities →

Recalled Product

Power Over Fiber Base Module

Share this recall: Facebook X / Twitter WhatsApp Email

Hazard / Issue

Device not certified - no laser product report was filed with CDRH prior to distribution. Product label incorrectly states device is "Class IV", although Class 1 is appropriate for device. Improperly labeled - omitted labels required under 21 CFR 1040.10 / IEC 60825-1 via LN#56. Missing certification and proper manufacturer's ID label. Was not accompanied by compliant User Information. Formalized quality control procedures and record keeping of QC testing was not performed to IEC 60825-1 nor 21 CFR 1040.10 standards. Sales information did not convey required class designation and warning statement

Issued by

FDA

Distribution: Nationwide (United States)
Lot/Code Info: Power Over Fiber Base Module Identification P/N 633232 P/N 622365 P/N 633383 P/N 633910 P/N 635731

✅ What Should You Do?

  1. 1Contact your healthcare provider immediately to discuss this recall.
  2. 2Stop using the device if it is safe to do so, and contact your provider.
  3. 3Follow the manufacturer's remedy instructions provided in the recall notice.
  4. 4Report device problems to FDA MedWatch or call 1-800-FDA-1088.

Looking for a safe alternative?

Search Amazon for Power Over Fiber Base →
View official government recall